Home/Brands/ ISOTECH/Industrial Calibration

ISOTECH Oceanus-6

Brand : ISOTECH
Categories : Industrial Calibration

The Oceanus-6 is a complete calibration laboratory working over the temperature range 45°C below ambient temperature to +140°C. It permits the calibration of temperature sensors absolutely (at fixed points of the ITS-90 scale) or by comparison to a reference standard.

  • Use with Water and Gallium Fixed Points
  • Can also be used as a Dry Block, Liquid Bath
  • -25°C to 140°C

GET QUOTE

Specifications

  • 0.01:

    -19.99 to 99.99, 0.1 100.0 to 140.0 PC can display 0.01 across whole range with the software included.

  • Model

    Oceanus-6

  • Power

    108 to 130V (50/60Hz) or 208-240V 950/60Hz) 300 Watts

  • Weight

    17kg

  • Cools from:

    20°C to -10°C in 90 minutes

  • Dimensions:

    Height 430mm Width 310mm Depth 300mm

  • Heats from:

    019°C to 80°C in 60 minutes

  • Dry Block Bath:

    ±0.03°C

  • Ice/Water Bath:

    ±0.001°C

  • Standard Insert

    6 pockets, all 8.0mm diameter and 250mm deep

  • Indicator Units:

    °C, °F, K

  • BlackBody Source:

    ±0.3°C

  • Temperature Range

    45°C below ambient to + 140°C

  • Computer Interface

    Included with software

  • Absolute Stability:

    (30 Minutes)

  • Calibration Volume:

    50mm diameter by 300mm deep

  • Display Resolution:

  • ITS-90 Fixed Point:

    ±0.0002°C

  • Stirred Liquid Bath:

    ±0.025°C

  • Surface Sensor Calibrator:

    ±0.5°C

The Oceanus-6 PLUS has all the advantages of the Isocal-6 models but with a substantially larger calibration volume, 50mm diameter by 300mm deep. The Oceanus-6 can be used as a Dry Block, a Liquid Bath, a Blackbody Source for infrared thermometers, a Surface Sensor Calibrator and for performance to a few mK (0.001°C) ITS-90 Fixed Points. The Oceanus-6PLUS is available in two models, the BASIC (B) and the SITE (S). The B model includes a sophisticated temperature controller with a dual display for Set Temperature and Dry Block Temperature.

The S model includes a built in digital thermometer to which an external standard thermometer can be connected, for Dry Block use this will give greater accuracy eliminating temperature gradient and loading errors.

For Liquid Bath, Surface Sensor or Blackbody use an external thermometer should always be used – either with the S model or the B model and a separate stand-alone indicator. For Lab use the Oceanus-6PLUS can be used with a laboratory performance temperature indicator such as one of the Isotech True Temperature Indicators with performance, for similar sensors, down to hundredths of a degree. The Oceanus-6PLUS offers unprecedented accuracies of ±0.0002°C (2 Sigma) at the Water triple point and the Gallium melt temperature of 29.7646°C and up to ±0.005°C in the stirred liquid bath option (by comparison).