Specifications

  • Model

    520

  • Power

    300 Watts, 108-130Vac or 208-240Vac, 50/60Hz

  • Dimensions

    Height 322mm, Width 176mm, Depth 263mm, Weight 14kg

  • Uniformity

    ±0.018°C

  • How to order

    520 Europa-6 POTTS. Please specify voltage needed.

  • Communications

    Supplied as standard with serial interface, PC adaptor cable and Cal Notepad.

  • Ice/Water Bath

    ±0.001°C

  • Heating/cooling

    See graph opposite

  • Indicator Units

    °C, °F, K

  • Metal Block Bath

    ±0.03°C

  • Black Body Source

    ±0.3°C

  • Temperature Range

    65°C below ambient to +140°C

  • Calibration volume

    35mm diameter x 160mm deep

  • ITS-90 Fixed Point

    ±0.001°C

  • Stabilisation Times

    10 minutes

  • Stirred Liquid Bath

    ±0.025°C

  • Indicator Resolution

    0.1 to 0.01 (4 digit display)

  • Absolute Stabilities:

    over 30 minutes:

  • Controller Resolution

    0.1 to 0.01 (4 digit display)

  • Surface Sensor Calibrator

    ±0.5°C

  • Absolute Minimum Temperature

    -45°C

The multi-functions of the Europa-6 is a unique award winning concept from Isotech.

The Europa-6 is a complete calibration laboratory working over the temperature range 65°C below ambient temperature to +140°C. It permits the calibration of temperature sensors absolutely (at fixed points of the ITS-90 scale) or by comparison to a reference standard.

It will calibrate both contact and non-contact thermometers such as optical pyrometers, surface sensors, as well as thermistors, thermocouples and resistance thermometers, whether they are short, long or odd shaped. Further it permits maintenance of reference standards by confirming the Water Triple Point at regular intervals.

The Europa-6 has a calibration volume of 35mm diameter by 160mm deep and is supplied with the very latest technology digital indicator and controller making the Europa-6 a complete self-contained calibration laboratory.

The Europa-6 offers unprecedented accuracies of 0.001°C (2 Sigma) at the Water Triple Point and the Gallium melt temperature of 29.7646°C and the Mercury Triple Point of -38.8344°C.