DESCRIPTION

Isotech has a wide range of Dry Blocks to suit probes requiring deeper immersion. This dry block temperature calibrator features large and deep calibration volumes. As such, it is less portable than earlier Dry Blocks, but has a higher capacity and retains outstanding temperature uniformity.

  • High Capacity Deep Block
  • Use for Comparison and Fixed Point Calibration
  • Use with very long thermometers

GET QUOTE

Specifications

  • Model

    510 Medusa 511 Medusa 3

  • Power

    108-130V or 208-240V 50/60Hz. 510: 1000 Watts 511: 1800 Watts

  • Cools from

    550°C to 30°C in 5 hours

  • Heats from

    30°C to 550°C in 90 minutes

  • How to order

    510 Medusa or 511 Medusa 3. Please specify model type, voltage and options required.

  • Uncertainties

    Refer to uncertainties graph

  • Indicator Units

    °C, °F, K

  • Standard Insert

    Six 8mm pockets all 250mm deep

  • BlackBody Source

    ±0.1°C

  • Metal Block Bath

    ±0.03°C

  • Temperature Range

    30°C to 550°C (510) and 50°C to 700°C (511)

  • Calibration Volume

    45mm diameter by 285mm deep

  • Computer Interface

    Included with software

  • Display Resolution

    0.01 to 99.99, 0.1 to 100.0 to 650.0. PC can display 0.1 across whole range with the software included.

  • ITS-90 Fixed Point

    ±0.001°C

  • Overall Dimensions

    Height 430mm, Width 310mm, Depth 300mm, Weight 510: 17kg 511:25kg

  • Absolute Stability:

    over 30 minutes:

  • Surface Sensor Calibrator

    ±0.5°C

Isotech have a wide range of Dry Blocks to suit probes requiring a large immersion depth. These products feature large and deep calibration volumes. As such they are less portable than the earlier Dry Blocks, but have higher capacities and retain outstanding temperature uniformity, this uniformity is so good that these larger products are also apparatus for Secondary Laboratories to realize the Fixed Points of ITS-90.

Medusa 510 has a maximum operating temperature of 550°C. In addition to the main heating zone there are additional top and bottom heaters which compensate for the end losses creating a constant temperature zone across the well.

For Comparison Calibration the Medusa should be used with an insert, the standard insert has six 8mm pockets 250mm deep. Also available is an insert 44mm diameter x 170mm deep which is suspended from the top of the block so that the height is user adjustable. For flexibility the Medusa can also be used with accessories for infrared thermometers and surface sensors. The Medusa is available in two models, the BASIC (B) and the SITE (S). The B model includes a sophisticated temperature controller with a dual display for Set Temperature and Dry Block Temperature.

The S model includes a built-in digital thermometer to which an external standard thermometer can be connected giving greater accuracy, eliminating temperature gradient and loading errors. Also included in the site model is a timer which can set the bath between two temperatures, and automate ITS-90 fixed point operation. For Surface Sensor and Blackbody use an external thermometer is recommended. For laboratory accuracy the Medusa can be used with a high-end temperature indicator such as an Isotech TTI model.

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature. Increased resolution of ±0.01 available throughout the range via the PC interface and from 0.01 to +99.99 locally on the autoranging front display. The controller features multi-point block to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers, Thermocouples (types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including 4-20mA current transmitters to be displayed on the in-built indicator. The indicator can be programmed with up to five calibration points to provide high accuracy digital probe matching. The indicator and controller are both addressable over the communications link.